HPLC Tools

Everyday liquid-chromatography calculations, in the browser. Nothing is uploaded; all the math runs locally, and your inputs are remembered on this device.

Method transfer & column scaling

Geometric scaling between columns (e.g. HPLC → UHPLC), including the gradient table and dwell-volume compensation, with checks against the USP <621> allowances.

Original method
New column / system

Gradient table

Enter the original program. Times are scaled to keep the same gradient volume per column volume. Leave a single row for an isocratic method.

Original time (min)%BRemove
New time (min)%BFlow (mL/min)
Equations used
  • Flow: F2 = F1 · (dc2²/dc1²) · (dp1/dp2). The dp term is optional.
  • Injection volume: V2 = V1 · (L2 dc2²)/(L1 dc1²)
  • Gradient times: t2 = t1 · (F1/F2) · (L2 dc2²)/(L1 dc1²). F2 includes any extra flow adjustment, so the gradient slope per column volume stays the same.
  • Pressure ratio (same mobile phase and temperature): P2/P1 = (F2/F1)(L2/L1)(dc1/dc2)²(dp1/dp2)²
  • Dwell: the original dwell volume is scaled to the new column volume and compared with the new system's dwell volume. A shortfall is made up with an initial isocratic hold. An excess can be offset by delaying injection, if the system supports it.

USP <621> (harmonized; official 1 Dec 2022) adjustment summary, based on USP's notice of the harmonized chapter:

  • L/dp: −25% to +50% of the original, for isocratic and gradient methods. When switching between fully and superficially porous particles, the efficiency criterion is used instead: N (isocratic) or (tR/w½)² (gradient) within −25% to +50%.
  • Flow: scaled with dc² and dp. A further ±50% is permitted.
  • Column temperature: ±10 °C (isocratic), ±5 °C (gradient). Mobile-phase pH: ±0.2 units. Minor components: ±30% relative or ±10% absolute.
  • Injection volume: may be adjusted, provided system suitability is met.

Always confirm against the current official chapter and the monograph, and verify system suitability after any change.